0102030405
High and Low Temperature Chamber | Reliable China Suppliers & Factory for Environmental Testing Solutions
Main Features and Advantages
-
1Integrated DesignCombines compressor refrigeration, electric heating, circulation pump, expansion tank, and control system into a single unit, saving space and simplifying installation.
-
2Wide Temperature RangeTypical temperature ranges span from -80°C to +300°C (or even higher), covering the majority of experimental and industrial needs. Specific ranges depend on the model and heat transfer fluid.
-
3High-Precision Temperature ControlUtilizes PID intelligent control, with temperature stability reaching ±0.1°C or even ±0.01°C, meeting the requirements of precise experiments.
-
4Fully Enclosed CirculationThe medium circulates in a sealed system, avoiding contact with air, which reduces oxidation, evaporation, and contamination, ensuring high safety and saving fluid.
-
5High Efficiency and SpeedPowerful heating and cooling capacities enable rapid achievement of target temperatures, shortening process waiting times.
-
6VersatilityA single unit can replace traditional setups such as "oil bath + low-temperature coolant circulation pump," enabling automated and continuous temperature ramp processes.
Core Application Areas
🧪 Chemistry and Pharmaceuticals
- Temperature control for reactors and glass reaction vessels.
- Temperature management during synthesis, crystallization, extraction, and distillation processes.
- Pilot-scale and industrial-scale production.
🔬 Biotechnology and Laboratories
- Temperature regulation for fermenters and bioreactors.
- Applications in cell culture, enzymatic reactions, and as a thermal source for PCR instruments.
- Cooling/heating support for rotary evaporators, cold traps, and freeze dryers.
⚡ Materials Testing and New Energy
- Environmental simulation for battery testing (e.g., lithium-ion batteries, fuel cells), including high-low temperature cycling tests.
- Testing of semiconductor components and chips.
- Research and performance testing of new materials (e.g., polymer materials).
🚗 Automotive and Aerospace
- Environmental simulation testing for automotive components (e.g., motors, battery packs, sensors).
- Testing of lubricants and fuels.
🏭 Petrochemical Industry
- Temperature control for analytical instruments such as viscometers and rheometers.
Selection Guide — Key Parameters to Consider When Purchasing
1 Temperature Range
Select based on the maximum and minimum temperature requirements of your experiment or process, with some allowance for margin.
2 Temperature Control Accuracy and Stability
Higher precision requirements typically correspond to higher equipment costs.
3 Heating and Cooling Capacity
The power rating determines the rate of heating and cooling. This should be calculated based on the volume and material of the reactor, as well as the desired temperature ramp rate.
4 Pump Flow Rate and Head
- Flow Rate (L/min): Determines the circulation speed and affects heat exchange efficiency.
- Head (m): Determines the ability to overcome pipeline resistance, especially when the reactor is elevated or the piping is long.
5 Heat Transfer Fluid
- Silicone Oil: Most commonly used, wide range (-40°C to 200°C), stable performance.
- Water/Ethylene Glycol Solution: Suitable for temperatures above 0°C, environmentally friendly and clean.
- Special High-Temperature Oil: For use above 300°C.
- Special Low-Temperature Fluid: For use below -80°C.
6 Internal vs. External Circulation
- Internal Circulation: Circulates the fluid within the device itself, primarily for maintaining the temperature of the internal bath.
- External Circulation: Connects to external equipment for fluid circulation. Ensure compatibility of interface dimensions and materials during selection.
7 Safety and Protection Features
Over-temperature protection, low-liquid-level protection, over-current protection, leakage protection, and self-diagnostic functions.
8 Programming Capabilities
Determine whether multi-segment temperature programming is needed for automated cyclic testing involving heating, holding, and cooling stages.
Typical Industrial Applications
Flowchart
Specification
| Parameter | UNIT | MODEL | ||||
| SOS-05 | SOS-10 | SOS-20 | SOS-30 | SOS-40 | ||
Products Detail
Frequently Asked Questions (FAQ)
Q What is the typical temperature range of a dynamic temperature control system?
Dynamic temperature control systems typically cover a wide range from -80°C to +300°C, depending on the model and heat transfer fluid selected. Low-temperature models can reach as low as -80°C, while high-temperature models can operate up to 300°C or beyond with special thermal fluids.
Q What heat transfer fluids are compatible with these systems?
Common heat transfer fluids include silicone oil (suitable for -40°C to 200°C), water/ethylene glycol solutions (for temperatures above 0°C), special high-temperature oils (above 300°C), and specialized low-temperature fluids (below -80°C). The choice of fluid depends on the required temperature range and application environment.
Q How accurate is the temperature control of these units?
These systems utilize PID intelligent control algorithms, achieving temperature stability of ±0.1°C in standard configurations and as precise as ±0.01°C in high-precision models. This level of accuracy is suitable for demanding laboratory and industrial applications.
Q What is the difference between internal and external circulation modes?
Internal circulation maintains the temperature of the fluid within the unit's own bath, while external circulation connects to outside equipment — such as reactors, fermenters, or test chambers — to regulate their temperature. Most industrial applications rely on external circulation. When selecting, ensure the interface dimensions and materials are compatible with your external equipment.
Q What safety protection features are included?
These systems are equipped with multiple safety protections, including over-temperature protection, low-liquid-level protection, over-current protection, leakage protection, and self-diagnostic functions. These features ensure safe and reliable long-term operation in both laboratory and industrial environments.
Q Can the system be programmed for automated temperature cycling?
Yes. Many models support multi-segment temperature programming, allowing users to set automated sequences of heating, holding, and cooling stages. This is especially useful for battery cycling tests, material aging studies, and pharmaceutical synthesis processes that require precise and repeatable thermal profiles.